For example when flow meter full scale is 100 L/min and accuracy is stated as +/- 5 % F.S. it means +/- 5 L/min accuracy on the flow meter whole measurement range.
% of F.S. inaccuracy is calculated
Flow meter reading – Real flow *100 = F.S. inaccuracy
Flow meter full scale
% RD accuracy
For example when flow meter reading is 100 L/min and accuracy is stated as +/- 5 % RD it means +/- 5 L/min accuracy. Or at 50 L/min reading +/- 2,5 L/min accuracy.
% of RD. inaccuracy is calculated
Flow meter reading – Real flow *100 = RD inaccuracy
Flow meter reading